Acoustic Velocity Transducers

ystcha 2018. 1. 24. 10:53

Acoustic Velocity Transducers

영신컨설턴트 (02) 529 8803 2018.1


Product Code : GDSAV

Systems to measure the P- and S- Wave velocities within a sample. AV sensors are generally used where pressures and load exceed those where bender elements can be used.

AV sensors are based on the same principle as bender elements but the piezo ceramic elements are not exposed to the environment so they can be used at pressures up to 100MPa and loads up to 2MN. Sensors and sources are mounted in the pedestal and topcap or to the sides of the sample in some cases. Each sensor package contains a Compressional wave (P-Wave) element and two shear wave (S-Wave) elements. The Shear wave elements are set in orthogonal directions to allow two shear waves to be generated with different polarisation. This is important where samples may be cross-anisotropic or even fully anisotropic.

 


Acoustic Velocity Transducer 암석에 적용 (구속압 100MPa)



bender element시험은 bender element가 지반 속에 삽입되어 지반에 적용되고 Acoustic Velocity Transducer는 삽입되지 않아 강도가 큰 시료에 적용한다.


bender element 지반에 적용



Acoustic Velocity Transducer 사양


Key features

Benefits to the user

Multiple Wave types as standard: 

 

Standard sensor package has: P-Wave, S- Wave Polarity 1 & S-Wave Polarity 2.

 

High speed data acquisition to yield high resolution results:

 

Result accuracy is dependent on the resolution of time in acquired data. Higher speed data acquisition yields higher resolution results.

 

High Bit count:

 

Yields a more dynamic capture of acquired waveforms with less tuning needed.

 

Designed to fit most GDS high pressure triaxial and Hoek Cells:

 

Some existing GDS high pressure cells can be adapted to include GDSAV transducer sets.  Please consult GDS for further information.

 

Vertical and horizontally mounted elements are available in some cells:

 

Allows users to describe fully anisotropic samples by measuring velocities in all directions and polarities. 

 

Software controlled switching:

Once the system is set up sensors do not have to be plugged and unplugged to use different wave types


Axial Load

Up to 2 on request

Data Acquisition

16 (Typically 12/14 at high logging rates)

Pressure Range (MPa)

Up to 100

Sample Sizes

38, 50 options available upon request

Weight Approx (kg)

Sample set 3.3 kg, Pedestal 0.95 kg & Top Cap 1.4 kg


Acoustic Velocity TransducerInstrumented Hoek Cell의 상하단에 설치



Acoustic Velocity Transducer 결과